| Catalog Number |
09180443001 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation of a software malfunction in the cobas infinity lab core license 3.X.The reporter stated that the instrument driver was saving a result in the field "r4" instead of the field "r4-1"; the driver was saving "62.42^" instead of "62.42".There was no report of any patient harm because of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The software version is 3.03.12.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined that the event occurred because the workflow manager driver was not able to capture the information as defined in the host interface manual of the instrument.The investigation verified that the event was a software issue in the cobas infinity.This issue has been addressed.
|
| |
|