| Model Number |
INTEGRON IT26-C |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No device-related deaths or serious injuries were informed.The malfunction is related to the absence of the pill in those particular chemical integrators.This affects the use of the product and cause the device to fail to perform its essential function and compromise the sterilization process monitoring effectiveness, as it may delay the release of a sterilization load, which could contribute to death or serious injury if it were to recur.
|
| |
|
Event or Problem Description
|
|
It was reported that two strips that didn't react to the steam.The use of the product was discontinuated.
|
| |
|