• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 30MM CENTRAL SCREW, MODULAR; SHOULDR PROSTH, REVERSE CONFIG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 30MM CENTRAL SCREW, MODULAR; SHOULDR PROSTH, REVERSE CONFIG Back to Search Results
Model Number 30MM CENTRAL SCREW, MODULAR
Medical Device Problem Codes Positioning Failure (1158); Compatibility Problem (2960)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 12/14/2023, it was reported by a sales representative via e-mail that an ar-9561-30s univers revers¿ modular glenoid system had an issue marrying the central screw to the ar-9560-24-2 base plate.This occurred during a reverse total shoulder on (b)(6) 2023, where the central screw did not marry to the base plate, and the metal ended up getting a little deformed.They followed all the correct steps from the technique guide and made sure the screw married the base plate correctly before tightening down with the clamp.No additional information was provided, and additional information has been asked.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The complaint allegation was confirmed.One unpackaged ar-9561-30s batch, 15100631, was received for investigation.Functional testing with the returned mating part ar-9560-24-2 batch 15097595 was not performed due to the damage to the screw hex; the base plate could get stuck.The visual evaluation found that the hex was damaged, with nicks, a possible sign of excessive force to screw/unscrew it from the base plate.The most likely cause of the reported failure is user error.When two devices are intended to be threaded together, ensure that they are fully engaged prior to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30MM CENTRAL SCREW, MODULAR
Common Device Name
SHOULDR PROSTH, REVERSE CONFIG
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18477995
Report Number1220246-2024-00224
Device Sequence Number12361854
Product Code PHX
UDI-Device Identifier00888867296572
UDI-Public00888867296572
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K173900
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number30MM CENTRAL SCREW, MODULAR
Device Catalogue NumberAR-9561-30S
Device Lot Number15100631
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/14/2023
Supplement Date Received by Manufacturer07/29/2024
Initial Report FDA Received Date01/09/2024
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-