• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERRAGENE S.A. STRYKER SRBI30; BIOLOGICAL INDICATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERRAGENE S.A. STRYKER SRBI30; BIOLOGICAL INDICATOR Back to Search Results
Model Number STRYKER SRBI30
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No device-related death or serious injury was reported.Investigation shows there is no malfunction of the auto-reader.The positive readout may have been caused by an ineffective sterilization process.However, although no information of the biological indicator was gathered, there's a possibility of bi malfunction, which may cause the device to fail to perform its essential function and compromise the sterilization process monitoring effectiveness which could contribute to delay patient treatment and, hence, to a serious injury.
 
Event or Problem Description
Sales representative reported malfunctioning of the autoreader incubator because of positive results obtained with the biological indicator after sterilization process in the vp4 equipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRYKER SRBI30
Common Device Name
BIOLOGICAL INDICATOR
Manufacturer (Section D)
TERRAGENE S.A.
ruta nac. no. 9, km 280
parque industrial micropi
alvear, santa fe 2130
AR  2130
Manufacturer (Section G)
TERRAGENE S.A.
ruta nac. no. 9, km 280
parque industrial micropi
alvear, santa fe 2130
AR   2130
Manufacturer Contact
hernando carrizo
ruta nac. no. 9, km 280
parque industrial micropi
alvear, santa fe 2130
AR   2130
MDR Report Key18478140
Report Number3013145340-2024-00010
Device Sequence Number12460989
Product Code FRC
UDI-Device Identifier07798164678793
UDI-Public07798164678793
Combination Product (Y/N)N
PMA/510(K) Number
K191021
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSTRYKER SRBI30
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2022
Initial Report FDA Received Date01/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-