| Model Number |
D400 |
| Medical Device Problem Codes |
High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Electric Shock (2554)
|
| Date of Event |
04/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this implantable cardioverter defibrillator (icd) received a commanded 41j shock due to high out of range shock impedance measurements.After the electric shock was given, the device displayed a code 1005, indicating an open circuit condition.Boston scientific technical services (ts) was consulted and recommended a right ventricular (rv) lead revision.No additional adverse patient effects were reported.At this time, this device system remains in service.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with this implantable cardioverter defibrillator (icd) received a commanded 41j shock due to high out of range shock impedance measurements.After the electric shock was given, the device displayed a code 1005, indicating an open circuit condition.Boston scientific technical services (ts) was consulted and recommended a right ventricular (rv) lead revision.Further information was received that a rv revision procedure was performed and it was surgically abandoned.The physician opted to replace this device due to therapy upgrade.No additional adverse patient effects were reported.
|
| |
|