U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION REZUM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION REZUM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number G2200
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a water vapor therapy procedure for benign prostatic hyperplasia, the generator displayed an error code 480 and a message that states: retract needle and remove the delivery device from patient, power off/on the generator if problem persists, which asked to retract the needle and remove the device, but the delivery device was not even attached or connected.The generator was unplugged and plugged back but the issue persisted.There were no patient complications.This event is being reported for an aborted/cancelled procedure with a patient under general anesthesia.
 
Event or Problem Description
It was reported that during a water vapor therapy procedure for benign prostatic hyperplasia, the generator displayed an error code 480 and a message that states: retract needle and remove the delivery device from patient, power off/on the generator if problem persists., which asked to retract the needle and remove the device, but the delivery device was not even attached or connected.The generator was unplugged and plugged back but the issue persisted.There were no patient complications.This event is being reported for an aborted/cancelled procedure with a patient under general anesthesia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REZUM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
NORTECH SYSTEM INC
nw 7791
1450
minneapolis MN 55485 7791
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18480486
Report Number2124215-2024-00683
Device Sequence Number6408326
Product Code GEI
UDI-Device Identifier08714729986850
UDI-Public08714729986850
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K180237
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG2200
Device Catalogue NumberG2200
Device Lot Number0000003650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/22/2023
Supplement Date Received by Manufacturer01/15/2024
Initial Report FDA Received Date01/09/2024
Supplement Report FDA Received Date01/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-