| Model Number |
4-840120DIJJ-JP |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while in use on a patient, the pb840 ventilator generated a background check alert diagnostic code and ventilation stopped.The patient was removed from the ventilator and was placed on an alternate ventilator without injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Section e: initial reporter cannot be provided due to japanese privacy regulations.Section e: japan medtronic personnel reported event to manufacturer on behalf of the customer.H3: medtronic has not received the suspect device/component from the customer for evaluation nor has the device been evaluated by the service engineer.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9 updated due to device evaluation section h6 evaluation codes updated due to device evaluation method - remove b17 and add b01 and b24 result - remove c20 and add c13 conclusion - remove d15 and add d02 component - remove g07003 and add g02005 h3: device evaluation summary: medtronic conducted an investigation based upon all information received.It was reported that while in use on a patient, the pb840 ventilator generated a background check alert diagnostic code and ventilation stopped.The device was available for evaluation.The service personnel (sp) inspected the unit, confirmed the reported issue with code "breath delivery (bd) electronically erasable programmable read-only memory (eeprom) checksum but could not reproduce the event.Sp suspected the possible cause to be faulty bd central processing unit (cpu) printed circuit board (pcb).Additionally, the sp found other parts unrelated to issue that also need to be replaced.Unit is awaiting customer approval to repair.The cause of the event isolated to potential fault with bd cpu pcba.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality specifications.The event is included in a trending and monitoring plan.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|