• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY CRT-D; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY CRT-D; NO MATCH Back to Search Results
Model Number CD3231-40
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient presented for implant procedure.During procedure it was noted that the setscrew fell off the pacemaker.The procedure was completed using a different pacemaker.The patient was in stable condition.
 
Additional Manufacturer Narrative
The reported field event of set screw anomaly was confirmed.Final analysis found that the right ventricular setscrew was missing, and the rv septum was damaged upon received.This is consistent with the reported field event of setscrew fell out during operation.The rv septum was damaged during operation could cause the setscrew fell out when it was over loosened.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Common Device Name
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18480775
Report Number2017865-2024-00752
Device Sequence Number20481550
Product Code NIK
UDI-Device Identifier05414734504362
UDI-Public05414734504362
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model NumberCD3231-40
Device Lot NumberA000144715
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/21/2023
Supplement Date Received by Manufacturer01/22/2024
Initial Report FDA Received Date01/09/2024
Supplement Report FDA Received Date01/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-