• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAB
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 02/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.Sv 3.0b.Customer returned pump for an alleged motor error alarm on event date 02-feb-2018 pump passed the displacement test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test.Pump history download using thds was successful.However, a47 alarms found on the history file which caused corrupted data.Unable to confirm the reported event date due to the corrupted data.A47 alarms are due to the pump not having power for a long period of time.The following were noted during visual inspection: cracked belt clip slot, cracked battery tube threads, stained end cap sticker, stained address/serial number label and cracked reservoir tube lip.The test p-cap/reservoir does lock into place.No moisture damage found on the electronics, motor, battery tube and vibrator assembly noted.Pump passed require testing.No motor error alarm noted.Motor passed motor test."select patient information cannot be provided due to regional privacy regulations." medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
Information received by medtronic indicated that the customer reported a motor error during the rewind.Troubleshooting was performed.No harm requiring medical intervention was reported.The customer will discontinue using the insulin pump and the pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
lusine boyadzhyan
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
MDR Report Key18481889
Report Number2032227-2024-104788
Device Sequence Number9016710
Product Code OYC
UDI-Device Identifier00643169513822
UDI-Public(01)00643169513822
Combination Product (Y/N)N
PMA/510(K) Number
P150019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAB
Device Catalogue NumberMMT-523LNAB
Device Lot NumberA4523LNABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2023
Initial Report FDA Received Date01/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-