On (b)(6) 2023, the lay user/patient contacted lifescan (lfs) usa, alleging that his onetouch ultra mini meter was displaying an ¿error 5¿ message.This complaint was classified based on information obtained from the customer care agent (cca) during the initial call and on additional information obtained after customer care (cc) reviewed the call recording.The patient alleged that the error message began to appear on (b)(6) 2023, at an unspecified time.The patient did not report how he manages his diabetes, nor whether he made any changes in response to the alleged issue.The cca noticed during the call, between 4:43 pm and 5:11 pm, that the patient was not okay and he claimed that his ¿hands were shaking¿.The patient stated that he called his doctor who asked him to test, however the patient was not able to do so due to the alleged issue.The patient did not report any medical treatment.At the time of troubleshooting, the cca noted the subject meter was not being used for the first time and the issue was resolved during troubleshooting.The cca established that the error message was caused by the patient¿s testing technique.No further action was taken.This complaint is being reported because the patient claims he was unable to test his blood glucose due to the reported issue and reportedly developed symptoms suggestive of a serious injury adverse event after the alleged meter issue began.There is insufficient information to rule out the contribution of the subject meter to the event.
|
|
The reported issue has been investigated.Investigation concluded that the device performed as intended by displaying an error message, protecting the user from an erroneous result.Similar complaints for this issue were trended for the test strip lot.It was concluded that the number of complaints for the lot did not breach thresholds indicative of a systemic issue.In addition, similar complaints for this issue were trended including the reported meter.It was concluded that the number of complaints for the meter did not breach thresholds indicative of a systemic issue.
|