MAUDE Adverse Event Report: STRYKER GMBH TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X180MM X 130°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X180MM X 130°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 31301180S
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.H3 other text : device discarded by the hospital.
 
Event or Problem Description
As reported: "a neck of femur fracture was to be treated with a gamma short nail.During the procedure an issue arose inserting the set screw.Details are unclear but difficulty caused them to remove the set screw and open a spare.Case was then completed without further issues that were raised.".
 
Event or Problem Description
As reported: "a neck of femur fracture was to be treated with a gamma short nail.During the procedure an issue arose inserting the set screw.Details are unclear but difficulty caused them to remove the set screw and open a spare.Case was then completed without further issues that were raised.".
 
Additional Manufacturer Narrative
The reported event could be confirmed, although the device was not returned but an image was shared by the customer which confirms the alleged issue.The device was not returned for evaluation but an image of it was shared by the customer which shows signs of misalignment evident by deformed and flattened threads.The deformation mark around the tip region where the transition (square to taper) starts also indicates the same, as under proper insertion this area cannot get deformed.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based only on the available information and the image, the issue is deemed to be user related.There are signs of misalignment observed in the set screw which most likely led to inadequate insertion of it during the surgery.If any further information is provided, the complaint report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCHANTERIC NAIL KIT, TI GAMMA3® Ø11X180MM X 130°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18482768
Report Number0009610622-2024-00021
Device Sequence Number2747251
Product Code HSB
UDI-Device Identifier04546540375124
UDI-Public04546540375124
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K034002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number31301180S
Device Lot NumberK170CBD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/15/2023
Supplement Date Received by Manufacturer02/15/2024
Initial Report FDA Received Date01/10/2024
Supplement Report FDA Received Date03/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-