| Model Number |
8884721252E |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Event or Problem Description
|
|
The customer reported when the probe was inserted, the weighted tip came loose.The tip remained in the patient's sinus with no possibility of removal at the time.
|
| |
|
Additional Manufacturer Narrative
|
|
H4 device manufacture date was added.The device history record (dhr) review showed that manufacturing and inspection of the product was performed as per applicable procedures and validated process.Photos were provided showing the weighted tip was detached from the feeding tube.No sample was received for evaluation.The reported condition was confirmed through the photos.A walk through was performed of the manufacturing process and all process and controls were found properly followed.The exact root cause could not be determined with the available information.Based on the historical review, a corrective and preventative action (capa) was generated to further investigate the reported condition.This complaint will be used for tracking and trending purposes.
|
| |
|