| Model Number |
STRYKER SLCD12 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No device-related death or serious injury was reported.The positive readout may have been caused by an ineffective sterilization process.However, as no further information was provided, we consider there's a possibility of bi malfunction, which may cause the device to fail to perform its essential function and compromise the sterilization process monitoring effectiveness which could contribute to delay patient treatment.
|
| |
|
Event or Problem Description
|
|
It was reported that there were recurring bis following completed vp4 cycles.The vp4 did not register any error codes or cancel any loads.
|
| |
|