| Catalog Number |
362760 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® cpt¿ cell preparation tube with sodium citrate a stopper deformation was found by the user.No health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields have been updated with corrected and/or additional information.D9: device available for evaluation: yes.D9: returned to manufacturer on: 26-dec-2023.H.6.Investigation summary: bdj received one (1) sample and four (4) photos were provided for investigation.The photos were reviewed and the customer¿s indicated failure mode for defective stopper molding was observed.Additionally, the customer sample was evaluated by visual examination and the indicated failure mode for defective stopper molding with the incident lot was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode defective stopper molding.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® cpt¿ cell preparation tube with sodium citrate a stopper deformation was found by the user.No health impact or consequence reported.
|
| |
|
Search Alerts/Recalls
|