| Catalog Number |
MX5341LN |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Separation (1562); Separation Problem (4043)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: other; device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that during an injected scan, the white part of the "tap" detached, causing the iodine to spill onto the scan table.This caused a 3-4 minute delay in performing the diagnostic scan (hematoma on left lower limb).Adverse patient effects are unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Date returned to mfg 2/5/2024.One decontaminated device was returned.Visual inspection under the microscope confirmed that the plug and body were separated and was not assembled correctly.Functional testing was not performed.There is no assignable root cause for the plug and body not being assembled correctly.No corrective actions will be performed.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
| |
|