• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PERCUFLEX URINARY DIVERSION STENT; CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PERCUFLEX URINARY DIVERSION STENT; CATHETER, NEPHROSTOMY Back to Search Results
Model Number M0061603300
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable investigation results of stent detached or separated.Block h10: the returned percuflex urinary diversion stent was analyzed, and a magnification and visual evaluation noted that the coil was kinked, and the stent shaft was detached.Additionally, the detached part was not returned.No other problems with the device were noted.The reported event of stent kinked was confirmed.Taking all available information into consideration, most likely, the device meets all manufacturing specifications required and passed all the controls and inspections, no abnormalities were reported during the assembly process.It is probable that the problem was caused due to the interaction between the guidewire and the stent, such as the handling of the device could have contributed with the complaint event reported.Regarding the reported problem of stent shaft detached, based on the irregular shape of the detached tip, there is enough evidence to conclude that the device was cut.For that reason, the as analyzed cause code is unintended use error cause or contributed to event.Therefore, all compiled information on this investigation determines that the most probable cause is unintended use error cause or contributed to event since the interaction between the user and device, or sample, caused or contributed to the error.
 
Event or Problem Description
It was reported to boston scientific corporation that a percuflex urinary diversion stent was used during a procedure performed on (b)(6), 2023.During the procedure, the stent was found damaged.It was noted that the stent was kinked.The procedure was successfully completed with another percuflex urinary diversion stent.There were no patient complications reported as a result of this event.Investigation results revealed that the stent was detached/separated, therefore this event has been deemed an mdr reportable event.Please see block h10 for full investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUFLEX URINARY DIVERSION STENT
Common Device Name
CATHETER, NEPHROSTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
spencer IN 47460
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18489189
Report Number2124215-2023-75216
Device Sequence Number19749336
Product Code LJE
UDI-Device Identifier08714729821144
UDI-Public08714729821144
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K830803
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM0061603300
Device Catalogue Number160-330
Device Lot Number0032028827
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/18/2023
Initial Report FDA Received Date01/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-