According to the reporter, during testing/maintenance, there was power handle error. the reboot failed and handle would not turn on.Charger bay, ports, cables all work well and no issues charging the handle.There was no patient involvement.Medtronic's initial evaluation of the incident device found that analysis of data stored on the handle shows evidence of field programmable gate array (fpga) reset/reprogram failures.
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the handle had 213 of 300 procedures remaining.Analysis of data stored on the handle shows evidence of field programmable gate array (fpga) reset/reprogram failures.According to this data, the handle was running v29.5 software at the time of the fpga reset/reprogram failures.It was reported that the handle shows evidence of field programmable gate array (fpga) reset/reprogram failures.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|