• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC EZ STEER¿ THERMOCOOL® BI-DIRECTIONAL CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC EZ STEER¿ THERMOCOOL® BI-DIRECTIONAL CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Catalog Number BNI75TCDFH
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 12/15/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a patient underwent a cardiac ablation procedure and patient experienced cardiac tamponade treated with a pericardiocentesis.Rf needle was not used for transseptal puncture.Ablation was performed.One hour after the procedure was started and during the ablation, impedance increased mid-ablation.The ablation was immediately stopped, but the blood pressure decreased.No steam pops have been confirmed.Pericardial drainage was performed and blood pressure was stabilized.The patient was conscious, but returned to the icu for monitoring.The patient's condition improved at this point.No error messages on equipment during procedure.The physician's opinion on the cause of the adverse event is the procedure.The physician mentioned that right ventricular outflow tract, the site of ablation, is normally indented, but the patient's right ventricular outflow tract was appeared even more indented than others on the computed tomography (ct) scan.The patient's breath was deep, so the tip of the catheter may have been stuck in the indented hollow during ablation, resulting in cardiac tamponade.The patient became stable and moved to general ward on the day after the procedure.Patient fully recovered.
 
Additional Manufacturer Narrative
E1 initial reporter phone: (b)(6).Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device number lot 31161468m and no internal action related to the complaint was found during the review.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ STEER¿ THERMOCOOL® BI-DIRECTIONAL CATHETER
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18489530
Report Number2029046-2024-00142
Device Sequence Number6991869
Product Code OAD
UDI-Device Identifier10846835003338
UDI-Public10846835003338
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P030031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberBNI75TCDFH
Device Lot Number31161468M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2023
Initial Report FDA Received Date01/10/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NGEN RF GENERATOR, JAPAN; UNK_CARTO 3; UNK_NGEN PUMP; UNSPECIFIED RF NEEDLE
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient SexFemale
-
-