SIGHT SCIENCES, INC. OMNI SURGICAL SYSTEM Back to Search Results
Catalog Number 1-108
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/13/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Physician successfully performed 180 degree canalplasty utilizing an omni device.Upon second advancement, the physician reporter the microcatheter buckled and snapped.The broken part of the microcatheter was removed from the canal.
 
Additional Manufacturer Narrative
The returned device was evaluated and was determined that there is no evidence that the device design or manufacturing process was the cause of the incident based on a review of the lot history and relevant components of the returned device.A video of the actual procedure was not available and therefore a definitive root cause could not be determined; however, the probable root cause for this incident is likely due to a surgical technique or use error based on: 1) returned device found to meet specification, and 2) user history of this physician as well as review of past similar complaints reported by other physicians, where videos of the procedure was provided, determined the incident occurred due to a surgical technique or use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNI SURGICAL SYSTEM
Common Device Name
OMNI SURGICAL SYSTEM
Manufacturer (Section D)
SIGHT SCIENCES, INC.
4040 campbell ave.
suite 100
menlo park CA 94025
Manufacturer (Section G)
SIGHT SCIENCES, INC.
4040 campbell ave
suite 100
menlo park CA 94025
Manufacturer Contact
rachel franco
4040 campbell ave.
suite 100
menlo park, CA 94025
MDR Report Key18490139
Report Number3010363671-2023-00027
Device Sequence Number3533562
Product Code MRH
UDI-Device Identifier00858027006525
UDI-Public00858027006525
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K202678
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Catalogue Number1-108
Device Lot NumberF23E11-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/13/2023
Supplement Date Received by Manufacturer12/13/2023
Initial Report FDA Received Date01/10/2024
Supplement Report FDA Received Date04/15/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-