• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P EMPOWR POROUS FEMUR WITH HA NANO, 7L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P EMPOWR POROUS FEMUR WITH HA NANO, 7L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS Back to Search Results
Catalog Number 244-01-107
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/12/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery, the agent reported "(revision knee on a djo press-fit knee)".The previous surgery and the surgery detailed in this event occurred 1 year and 1 month apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for review.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was an ncmr#61385 associated with the main part #244-01-107, empowr porous femur with ha nano, 7l which documents that out of 18 parts lot 1 part was rejected due to spot/carbon mark on flange area.Later, the rejected part was found with no defects and accepted after proper justification.All other items in the respective lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause for this complaint was a revision on a press-fit knee.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.
 
Event or Problem Description
Revision surgery - revision knee on a djo press-fit knee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR POROUS FEMUR WITH HA NANO, 7L
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758
MDR Report Key18490184
Report Number1644408-2023-01976
Device Sequence Number3665443
Product Code MBH
UDI-Device Identifier00190446364481
UDI-Public00190446364481
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K210308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number244-01-107
Device Lot Number3248A1008
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2023
Initial Report FDA Received Date01/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
341-11-707 LOT: 4011A1000; 353-01-107 LOT: 192W1027
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient SexMale
-
-