| Model Number |
EG-2990K |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The time of event is unknown.There was no report of patient harm.Air/water nozzle missing.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the air nozzle missing.Based on the result, we concluded that it was caused due to the physical damage applied on the air nozzle.In addition, our technician confirmed that the water nozzle missing; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.As a result of confirming the detail as gfe, we got a ""no"" answer to the question of whether the nozzle was fallen into the human body or not.Therefore, based on the technical report""hr-rpt-0585(nozzle)"" and/or the risk analysis results, it was evaluated not to submit mdr.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information.B4: date of this report.G6: follow up #1.H2:if follow-up, what type? there were typos in some of the investigation results, which have been corrected.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the air nozzle missing.Based on the result, we concluded that it was caused due to the physical damage applied on the air nozzle.In addition, our technician confirmed that the water nozzle missing; however, this defect is not the main cause, and/or irrelevant to the alleged complaint.In terms of the nozzle missing, the possibility of dropping into human body could not be denied.Therefore, based on the technical report ""hr-rpt-0585(nozzle)"" and/or the risk analysis results, it was evaluated to submit mdr.
|
| |
|
Search Alerts/Recalls
|