• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM; BALLOON EXPANDABLE VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM; BALLOON EXPANDABLE VASCULAR COVERED STENT Back to Search Results
Catalog Number LSMU1350637
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 07/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during the preparation of a stent graft placement procedure in the common iliac artery via brachy access, the mounted stent allegedly fell out when the protector was removed.The procedure was completed using another device.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTREAM
Common Device Name
BALLOON EXPANDABLE VASCULAR COVERED STENT
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI  N A
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18490626
Report Number9616666-2023-00273
Device Sequence Number15217932
Product Code PRL
UDI-Device Identifier05391522081062
UDI-Public(01)05391522081062
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P160024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberLSMU1350637
Device Lot NumberCMHV0378
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/20/2023
Initial Report FDA Received Date01/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-