• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NATUS MEDICAL INCORPORATED OBM DAB (DIGITAL ACQUISITION BOX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NATUS MEDICAL INCORPORATED OBM DAB (DIGITAL ACQUISITION BOX) Back to Search Results
Model Number OBM00002
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial report ref natus complaint (b)(4) the device was used with a patient for approximately 19 hours.The customer was given an adverse event questionnaire to complete and return for natus's investigation.The customer stated they will work with the clinical handler to review and provide answers.The customer was shipped 1 obm dab (digital acquisition box), serial number:(b)(6) for exchange.Risk file review: doc-013950 rev 11 olympic brainz monitor(obm) risk analysis spreadsheet (ras) hazard id - 6.13 - inadequate packaging / shipping specifications effects (harm): data loss.Delayed diagnosis.The hazards identified have been reduced as far as possible, and the residual risk for these hazards are categorized as having an rba rating of low due to the nature of the hazard.Hazards have associated warnings disclosed in the ifu.Risk outweighed by benefit of use of device.Further investigation to be carried out.
 
Event or Problem Description
Part obm00002 obm dab (digital acquisition box) mislabeled.The customer reported experiencing issues with their obm (olympic brains monitor) having the electrode placement sticker on backwards, resulting in the electrodes being applied incorrectly.
 
Additional Manufacturer Narrative
Follow up report 002 ref natus complaint#(b)(4) update to risk file review: per doc-013950 rev 12 olympic brainz monitor(obm) risk analysis spreadsheet (ras) hazard id - 5.39 - device label depicting electrode connection is attached in the wrong orientation.User misinterprets device label and connects electrodes so that the left and right channels are swaped effects (harm): wrong diagnosis with no clinical effects.The hazards identified have rba rating of negligible and are deemed broadly acceptable.These risks have been reviewed to determine if any further control measures may be implemented so as to further reduce the risk as far as possible.Risk outweighed by benefit of use of device.Device was returned 08 february 2024 13 february 2024 sustaining engineer provided evaluation details - operator error leading to label applied incorrectly at contract manufacturer creation.Scar-000067 created to document issue with the supplier.Work instructions qms-004738 shows correct orientation for applying label to the product.There is no additional documented downstream verification for correct label application.Label and cutout where label is to be applied is symmetrical in shape.This allows the label to fit properly into the cutout in both the correct and incorrect orientations.Failure confirmed: yes investigation result code: neuro sbu|user error.
 
Event or Problem Description
Part obm00002 obm dab (digital acquisition box) mislabeled.The customer reported experiencing issues with their obm (olympic brains monitor) having the electrode placement sticker on backwards, resulting in the electrodes being applied incorrectly.
 
Additional Manufacturer Narrative
Follow up report ref natus complaint# (b)(4).Jan 10, 2024 -technical service reached out to the customer in a second attempt to get them to complete the questionnaire and to return the defective device.Jan 15, 2024: customer reached out and reported the device was in fact not used with a patient.Customer stated they will not fill out and return their form.Customer confirmed they will return the data acquistion box.Feb 02, 2024 - awaiting return of the affected device for evaluation.
 
Event or Problem Description
Part obm00002 obm dab (digital acquisition box) mislabeled.The customer reported experiencing issues with their obm (olympic brains monitor) having the electrode placement sticker on backwards, resulting in the electrodes being applied incorrectly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OBM DAB (DIGITAL ACQUISITION BOX)
Common Device Name
OBM DAB (DIGITAL ACQUISITION BOX)
Manufacturer (Section D)
NATUS MEDICAL INCORPORATED
dba excel-tech ltd.
2568 bristol circle
oakville, ontarios, L6H 5 S1
CA  L6H 5S1
Manufacturer (Section G)
NATUS MANUFACTURING LIMITED
ida business park
gort
galway h91pd92,
EI  
Manufacturer Contact
gráinne walsh
dba excel-tech ltd.
2568 bristol circle
oakville, ontarios, L6H 5-S1
CA   L6H 5S1
MDR Report Key18490690
Report Number9612330-2024-00003
Device Sequence Number17060026
Product Code OMC
UDI-Device Identifier00382830010825
UDI-Public00382830010825
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K093949
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOBM00002
Device Catalogue NumberOBM00002
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/18/2023
Supplement Date Received by Manufacturer12/18/2023
12/18/2023
Initial Report FDA Received Date01/11/2024
Supplement Report FDA Received Date02/09/2024
03/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-