| Model Number |
SHEATH,HF,TAP/FEN,2 STPCK FOR 4MM SCOPE |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 12/15/2023, it was reported by a sales representative that an (b)(6) sheath seal was damaged and caused leaking.This was discovered during a procedure.The patient was not affected and there were no adverse events.The case was completed as normal, and the staff relied on the drape to collect the dripping arthroscopy fluid.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.Upon functioning testing, it was noticed that the device was leaking from the black button (accordion seal).Visual inspection revealed that the device exhibited signs of wear and tear.The most likely cause for the reported failure can be attributed to wear and tear.
|
| |
|