Model Number EQUINOXE CAGE GLENOID LARGE, BETA
Medical Device Problem Codes Fracture (1260); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 09/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section d10: concomitant products: equinoxe, humeral stem primary, press fit 13mm (cat#: 300-01-13 / serial#: (b)(6).Equinoxe, humeral head tall, 47mm (beta) (cat#: 310-02-47 / serial#: (b)(6).Equinoxe replicator plate 4.5mm o/s (cat#: 300-10-45 / serial#: (b)(6).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported by the equinoxe shoulder study, approximately two years post initial right tsa, the 67 y/o patient experienced cage liner dissociated.Patient had a revision on (b)(6)2024.Per clinical study report the reported event is definitely related to the device and the procedure.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reported glenoid failure may be the result of glenoid loosening and disassembly of the center peg from the polyethylene body.However, the failure and potential contributions from manufacturing, patient, or user-related issues to the event cannot be confirmed as the devices were not returned for evaluation and no images or radiographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE CAGE GLENOID LARGE, BETA
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
miguel soza
MDR Report Key18499028
Report Number1038671-2024-00082
Device Sequence Number7802168
Product Code KWT
UDI-Device Identifier10885862172716
UDI-Public10885862172716
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/13/2023
Device Model NumberEQUINOXE CAGE GLENOID LARGE, BETA
Device Catalogue Number314-13-14
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2023
Supplement Date Received by Manufacturer05/21/2025
Initial Report FDA Received Date01/11/2024
Supplement Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/15/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
Patient Weight98 KG
Patient RaceWhite
-
-