| Model Number |
EQUINOXE CAGE GLENOID LARGE, BETA |
| Medical Device Problem Codes |
Fracture (1260); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
09/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section d10: concomitant products: equinoxe, humeral stem primary, press fit 13mm (cat#: 300-01-13 / serial#: (b)(6).Equinoxe, humeral head tall, 47mm (beta) (cat#: 310-02-47 / serial#: (b)(6).Equinoxe replicator plate 4.5mm o/s (cat#: 300-10-45 / serial#: (b)(6).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported by the equinoxe shoulder study, approximately two years post initial right tsa, the 67 y/o patient experienced cage liner dissociated.Patient had a revision on (b)(6)2024.Per clinical study report the reported event is definitely related to the device and the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The reported glenoid failure may be the result of glenoid loosening and disassembly of the center peg from the polyethylene body.However, the failure and potential contributions from manufacturing, patient, or user-related issues to the event cannot be confirmed as the devices were not returned for evaluation and no images or radiographs were provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|