| Model Number |
100316 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device subject of the event was discarded by the distributor and not returned to medivators for evaluation.Without the returned device, a root cause could not be determined.The device history record was reviewed, and no abnormalities were noted.There have been no other complaints associated with this lot.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The distributor reported that a suction valve from their defendo valves & kits "can no longer be released/closed after opening" during a patient procedure resulting in suctioning of the mucous membrane in the stomach and bleeding.The user switched to a different valve to complete the procedure.No additional medical intervention was required.
|
| |
|