| Model Number |
328418 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Consumer reported that the rubber stopper is damaged, he stated that it appears to melted in the barrel.Consumer also reported that the plunger rod is difficult to move.Consumer does not re-use.Consumer only saved the syringe with the damaged stopper, the other syringes were discarded.Lot #: 3030357.Catalog #: 328418.Date of event: unknown.Samples: available - sending mail kit.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
| |
|