| Model Number |
48-30-00 |
| Medical Device Problem Codes |
No Audible Alarm (1019); No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the sorin s5 system.The incident occurred in germany.A livanova field service representative was dispatched to the facility to investigate the device.According to customer information, the issue is a sporadic error, no acoustic alarms (entire hlm), alarm only occurs visually.The alarm displayed was not reported and it is unknown.The console was tested, checked of all alarms (return after muting, high and low priority), plug connectors checked, recording carried out using pcan explorer, no error messages found on the can bus.The fault complained about could be resolved within 3 hours.Test run cannot be reproduced.Test run ok and no spare part was requested.Unit returned to customer.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova deutschland has received a report that, during priming, the s5 console displayed visual alarm and no acoustic alarm.Medical team switched the machine off and on and it worked for a while; then no acoustic alarm was heard again.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A livanova field service engineer (fse) was dispatched on-site and confirmed the reported issue.The fse troubleshot on-site and checked all the involved connections and was able to solve the issue without any parts replacement.Mechanical/electrical checks and functional testing were performed and passed successfully, therefore the unit was put back into regular service.The complaints database review pointed out that no further similar issue has been submitted for this unit since its installation in 2017.Based on the above facts, the root cause of the reported case is an electrical failure, as bad/loose connections, a false contact or some oxidized electronics that were reestablished and solved by the technician during service intervention.Missing alarm sound is not likely to result in serious injury or death since does not prevent visible alerts/messages on the pump/system panel display allowing user to realize the alarm situation and does not impair safety measures (e.G.Clamp closure, pump stop) and functionality of the s5 system.Therefore, the event has been reassessed as not reportable.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|