| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the images recorded/stored in mosaiq are not related to reference computerised tomography (ct)/ cone beam computed tomography (cbct).
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h6 updated section h11 updated the investigation was completed by conducting a thorough evaluation of the product and the reported information the user followed an incorrect workflow to re-ct the patient during treatment.To re-ct the patient, the user should have created a second prescription & site setup in mosaiq's d&i (diagnosis and interventions) for the new reference ct to be imported to the second site setup which maintained the original data allowing treatment to continue under the newest prescription which leaves the previous cbct still attached to the older site setup.The site setup can only have one image associated.The user overwrote the original image by importing a new site setup which caused this issue to be seen.The older images are no longer under the original site setup as the user overwrote those with the wrong workflow.The root cause was determined to be use error.Mosaiq did not have any malfunction and was working as designed and intended.There was no patient mistreatment.
|
| |
|