COM-DA HEALTHCARE CO.,LTD DRIVE DEVILBISS HEALTHCARE; CANE Back to Search Results
Model Number HCANE-BK-C2
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Codes Fall (1848); Laceration(s) (1946)
Date of Event 08/18/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Drive devilbiss healthcare was notified of a complaint involving a cane by an end user who stated, "the foot of his hurrycane came off causing him to slip and fall face first onto cement." the end went to urgent care where he received 8 stitches on his forehead.A replacement cane was sent to the end user.The hurrycane was returned to drive devilbiss for evaluation and was confirmed that the base of the hurrycane was detached.The screw that attaches the base to the cane appears to be unscrewed and loose, causing the base to be detached.The screw head had worn into the rubber base and broke surface which can cause stability issues.The hurricane that was returned reflects the design of the original hurrycane 1.0 where the base is detachable.Drive devilbiss no longer sells this version.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVE DEVILBISS HEALTHCARE
Common Device Name
CANE
Manufacturer (Section D)
COM-DA HEALTHCARE CO.,LTD
no. 1, jinzhao street, west
district, no. 6 workshop ii
zhongshan city, guangdong
CH 
Manufacturer (Section G)
COM-DA HEALTHCARE CO.,LTD
no. 1, jinzhao street, west
district, no. 6 workshop ii
zhongshan city, guangdong
CH  
Manufacturer Contact
kevin walls
33 golden eagle lane
littleton, CO 80127
7209625412
MDR Report Key18501274
Report Number3014944996-2024-00001
Device Sequence Number8125982
Product Code IPS
UDI-Device Identifier00822383571034
UDI-Public822383571034
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberHCANE-BK-C2
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2024
Initial Report FDA Received Date01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-