| Model Number |
M00514250 |
| Medical Device Problem Codes |
Break (1069); Human-Device Interface Problem (2949); Device Stenosis (4066)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
12/11/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an ultraflex esophageal proximal release covered stent was implanted in the esophagus for the treatment of a 2cm stenosis due to esophageal cancer during a gastroscopy procedure performed on (b)(6) 2023.The patient's anatomy was not tortuous and was not dilated prior to stent placement.On (b)(6) 2023, it was observed that the stent's distal end had overgrowth.The stent was moved slightly distally using rat tooth grasping forceps.During the procedure, it was also noted that the stent was obstructed by a small amount of food.The physician pushed past some impacted food and found that the stent had broken into two pieces.The patient was referred to a different hospital for stent removal and placement of a new stent.No patient complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of stent break.Imdrf device code a22 captures the reportable event of stent overgrowth.Imdrf device code a0106 captures the reportable event of stent obstruction.Imdrf patient code e2008 captures the reportable event of unretrieved device fragment.Imdrf impact code f23 captures the reportable event of additional intervention performed to adjust the position of the stent.
|
| |
|