| Catalog Number |
05964075001 |
| Medical Device Problem Codes |
Imprecision (1307); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The na electrode lot number and expiration date were not provided.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable gen.2 ise indirect for na results for 1 patient sample on a cobas 8000 cobas ise module.The initial na result was 136 mmol/l.The repeat result was 130 mmol/l.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The root cause of the event could not be determined.
|
| |
|