C.R. BARD, INC. (COVINGTON) -1018233 BARD® TIEMANN MODEL UROLOGICAL CATHETER COUDE TIP Back to Search Results
Catalog Number 010116
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that patient has been using intermittent catheter for several years, over the last two weeks caller and patient had been experiencing a problem during insertion.Caller stated during insertion catheter got to a certain point and they were met with resistance.Representative confirmed during insertion the hook was facing upward.Advised caller to contact patient urologist to schedule an appointment.Advised not to attempt to push against the catheter resistance as it could cause urethral irritation.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned for evaluation.A potential root cause for this failure could be "outside diameter is too large, rough or irregular shape protrusions".The dhr review could not be performed without a lot number.The instructions for use were found adequate and state the following: "instructions for use indications for use: for urological use only.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box caution: this product contains natural rubber latex which may cause allergic reactions.Warnings: on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions: urinary tract infection.Bleeding from the urethra.Irritation of the urethra.Instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands." section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that patient has been using intermittent catheter for several years, over the last two weeks caller and patient had been experiencing a problem during insertion.Caller stated during insertion catheter got to a certain point and they were met with resistance.Representative confirmed during insertion the hook was facing upward.Advised caller to contact patient urologist to schedule an appointment.Advised not to attempt to push against the catheter resistance as it could cause urethral irritation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® TIEMANN MODEL UROLOGICAL CATHETER COUDE TIP
Common Device Name
UROLOGICAL CATHETER COUDE TIP
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18503028
Report Number1018233-2024-00112
Device Sequence Number15485145
Product Code KOD
UDI-Device Identifier00801741029875
UDI-Public(01)00801741029875
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number010116
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer01/24/2024
Initial Report FDA Received Date01/12/2024
Supplement Report FDA Received Date02/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-