Model Number 1258T/75
Medical Device Problem Codes Capturing Problem (2891); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer reference number:2017865-2024-01018.It was reported that the patient presented in the clinic for a follow-up.Upon interrogation, the right ventricle (rv) lead exhibited noise that was oversensed by the device and led to pacing inhibition.Also, the left ventricular (lv) lead was observed with high capture threshold.Flouroscopy confirmed the lv lead had dislodged.Both rv and lv leads was capped and replaced on (b)(6) 2023.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Common Device Name
Drug eluting permanent left ventricular (lv) pacemaker electrode
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18504177
Report Number2017865-2024-01017
Device Sequence Number7864018
Product Code OJX
UDI-Device Identifier05414734502603
UDI-Public05414734502603
Combination Product (Y/N)Y
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2012
Device Model Number1258T/75
Device Lot Number3123649
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2023
Initial Report FDA Received Date01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/12/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PM3222 ALLURE.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexMale
-
-