• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1804225-28
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat the proximal to mid circumflex coronary artery with heavy calcification.A dissection was noted.The 3.0x28 mm xience skypoint stent delivery system (sds) failed to cross the lesion and was not long enough for the dissection, therefore the sds was removed and replaced with a 33 mm stent.The 2.25x28 mm xience skypoint sds was then attempted to be advanced but the physician did not want to cross the bifurcation due to the dissection.Upon removal from the anatomy, a proximal stent strut became stuck and the stent dislodged from the delivery system.A guidewire was attempted to be used to remove the stent but this was unsuccessful.The stent remains in the anatomy within the target lesion.The dissection was ultimately not treated but seemed to have resolved by the end of the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure was to treat the proximal to mid circumflex coronary artery with heavy calcification.A dissection was noted.The 3.0x28 mm xience skypoint stent delivery system (sds) failed to cross the lesion and was not long enough for the dissection, therefore the sds was removed and replaced with a 33 mm stent.The 2.25x28 mm xience skypoint sds was then advanced and was intended to treat the dissection but the physician did not want to cross the bifurcation due to the dissection.Upon attempted removal from the anatomy, a proximal stent strut became stuck and the stent dislodged from the delivery system.A guidewire was attempted to be used to remove the stent but this was unsuccessful.The stent remains in the anatomy within the target lesion.The dissection was ultimately not treated but seemed to have resolved by the end of the procedure.Subsequent to the initially filed report, it was reported that after the 3.0x28 mm xience skypoint sds failed to cross the lesion, a 33 mm xience stent was implanted.Next, the 2.25x28 mm xience skypoint sds reached the lesion but was removed as it was deemed a longer stent was needed.There was no resistance during advancement.The 2.25x28 mm xience skypoint sds caught on a proximal strut of a previously implanted xience 33 mm stent.There was no damage noted to the 33 mm xience stent.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties and treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.H6 - medical device problem code 2524 was removed.
 
Additional Manufacturer Narrative
Nasubsequent to filing the final report, it was determined that the medwatch received and attached to the previous report was not associated with this complaint; therefore, the medwatch is being removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18504532
Report Number2024168-2024-00720
Device Sequence Number20864299
Product Code NIQ
UDI-Device Identifier08717648233029
UDI-Public08717648233029
Combination Product (Y/N)Y
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/07/2025
Device Catalogue Number1804225-28
Device Lot Number2121541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/21/2023
Supplement Date Received by Manufacturer02/12/2024
02/27/2024
Initial Report FDA Received Date01/12/2024
Supplement Report FDA Received Date03/01/2024
03/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexMale
Patient Weight92 KG
-
-