| Catalog Number |
2C6891 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that thean interlink system extension set with control-a-flo regulator ruptured.This occurred during use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: during follow up, it was clarified that there was no issue with the baxter interlink system extension set.The rupture was caused by fluids being pushed through the dial-a-flow with too much pressure from another tubing that had a bulb on it.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|