• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMPL TAPERED SP 3.7MM 8MM OCTAGON; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMPL TAPERED SP 3.7MM 8MM OCTAGON; DENTAL IMPLANT Back to Search Results
Catalog Number SPB8
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Necrosis (1971); Pain (1994)
Date of Event 09/05/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant at tooth site 36 was removed due to necrotic tissue around the implant.Pain was reported as a result of the event.Patient had to return to place a new implant.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).G4: premarket identification k011245|k002188.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D4: additional device information.D9: device availability.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) spb8, (impl tapered sp 3.7mm 8mm octagon) for evaluation.Visual evaluation was performed, the implant had signs of use with markings from removal.No damage identified or signs of malfunction that would contribute to the event.Measurements match drawing.Dhr review was completed for the subject lot number 2023021215.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2023021215 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : medical : other.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was inadequate treatment planning.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TAPERED SP 3.7MM 8MM OCTAGON
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key18505108
Report Number0002023141-2024-00073
Device Sequence Number8145656
Product Code DZE
UDI-Device Identifier00889024017207
UDI-Public(01)00889024017207(17)280315(10)2023021215
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K082639
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 06/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberSPB8
Device Lot Number2023021215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/01/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date01/12/2024
Supplement Report FDA Received Date06/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/14/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-