It was reported to boston scientific corporation that a jagtome revolution rx was used in the sphincter of oddi during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6), 2024.During the sphincterotomy procedure, after the device was tensioned, the cutting wire of the jagtome revolution rx broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h10: the returned jagtome revolution rx was analyzed, and a visual evaluation noted that the cutting wire was broken, blackened and kinked from the proximal pierce hole, which are consistent with the findings when the device was observed under magnification.Additionally, the working length was torn at the green band at the rapid exchange channel.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, kinked, and blackened.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can bend the cutting wire.Additionally, the working length was torn from the brown band to the tip, which could have been caused by pulling/removing the guidewire through the c-channel, or the technique used during loading/removing the guidewire into the device.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
|
It was reported to boston scientific corporation that a jagtome revolution rx was used in the sphincter of oddi during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During the sphincterotomy procedure, after the device was tensioned, the cutting wire of the jagtome revolution rx broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|