Catalog Number 05342716001
Medical Device Problem Codes Break (1069); Fracture (1260); Material Integrity Problem (2978)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported an issue with the cover of the benchmark ultra stainer module.The operator was loading reagents and the cover reportedly fell onto the operator's wrist.
 
Additional Manufacturer Narrative
The reason the cover reportedly fell is being investigated.
 
Additional Manufacturer Narrative
The lid counterbalance and images of it were requested for investigation, but these could not be provided.The investigation determined the issue is consistent with overextension of the hinge during the use of the instrument.Mechanical stops on the hinge of the lid control how far the hinge will open (85 degrees plus or minus 5 degrees).Opening or forcing the lid past the typical 85 degrees increases the chances of the pin and connector bushing detaching, allowing the compression spring assembly to fall out of the hinge.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENCHMARK ULTRA
Common Device Name
SLIDE STAINER, AUTOMATED
Manufacturer (Section D)
VENTANA MEDICAL SYSTEMS INC.
1910 e innovation park dr
tucson AZ 85755
Manufacturer (Section G)
VENTANA MEDICAL SYSTEMS INC.
1910 e innovation park dr
tucson AZ 85755
Manufacturer Contact
timothy blair
1080 us hwy 202 s
branchburg, NJ 08876
MDR Report Key18507214
Report Number2028492-2024-00130
Device Sequence Number15428935
Product Code KPA
UDI-Device Identifier04015630981052
UDI-Public04015630981052
Combination Product (Y/N)Y
Initial Reporter CountryFR
PMA/510(K) Number
510K EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05342716001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/20/2023
Supplement Date Received by Manufacturer02/02/2024
Initial Report FDA Received Date01/12/2024
Supplement Report FDA Received Date02/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-