| Catalog Number |
05342716001 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260); Material Integrity Problem (2978)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported an issue with the cover of the benchmark ultra stainer module.The operator was loading reagents and the cover reportedly fell onto the operator's wrist.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason the cover reportedly fell is being investigated.
|
| |
|
Additional Manufacturer Narrative
|
|
The lid counterbalance and images of it were requested for investigation, but these could not be provided.The investigation determined the issue is consistent with overextension of the hinge during the use of the instrument.Mechanical stops on the hinge of the lid control how far the hinge will open (85 degrees plus or minus 5 degrees).Opening or forcing the lid past the typical 85 degrees increases the chances of the pin and connector bushing detaching, allowing the compression spring assembly to fall out of the hinge.
|
| |
|