| Model Number |
DFR00V |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Section: a4 and a5: unknown/ asked but not available.Section e1: surgeon email asked but not provided.Section h3: other 81: the device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the preloaded intraocular lens (iol) was removed and replaced from the left eye due chip or chunk of the lens was missing.The procedure was completed using lens of same model.The patient is doing good.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d-9: device available for evaluation: yes.Section d-9: device date returned to manufacturer: 26 jan 2024.Section h-3: device evaluated by manufacturer: yes.Device evaluation: visual inspection reveal that the lens was observed to be coated in viscoelastic residue.The lens was cleaned and inspected revealing that the lens received was not a johnson & johnson model lens.No further evaluation was performed on the lens.The simplicity was visually inspected revealing that the cartridge tip was damaged.No other damage to the complaint simplicity was observed manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order was performed.Conclusion: the complaint issue reported could not be confirmed and no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|