| Model Number |
VICM5 12.1 |
| Medical Device Problem Codes |
Optical Distortion (3000); Optical Problem (3001)
|
| Health Effect - Clinical Codes |
Visual Disturbances (2140); Halo (2227)
|
| Date of Event |
11/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.1mm vicm5_12.1 implantable collamer lens of -4.00 diopter was implanted into the patient's right eye (od) on (b)(6) 2023.On 30-nov-2023 the lens was exchanged for a longer lens of the same model and power due to glare/haloes and the problem resolved.Cause of event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: h6: medical device problem code: "3000" should be removed and "3001" should be added.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h6.- type of investigation code (b): 3331- device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
|
| |
|