| Model Number |
PE1115-1413 |
| Medical Device Problem Codes |
Nonstandard Device (1420); Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The surgeon reported that prior to implanting the ustar ii curved stem he noticed that the direction of bend was incorrect.He then implanted the ustar ii straight stem and completed the surgery without incident.No patient harm was reported.Examination of the device against the drawing confirmed that the stem is bent at a wrong angle, 180 degrees from that intended.Root cause investigation found that production does not have poka-yoke for the bending direction and that the testing process for identifying improper bending was inadequate.A recall of ustar ii curved stems is being conducted.Returned devices will be inspected, and the defective items destroyed.New devices will be produced and tested in accordance with the findings of the investigation.
|
| |
|
Event or Problem Description
|
|
During a hip reconstruction surgery in taiwan, the surgeon found that the hip stem curvature of the selected hip stem was incorrect prior to implanting the device and the surgeon ended up using a straight stem to finish the case.No patient harm was reported.
|
| |
|