| Catalog Number |
ES2914 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, non-conforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no non-conforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information when attempting to inflate the device it returned to flaccid state.The device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump, and cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted with the pump.No functional abnormalities were noted with either cylinder 1 or cylinder 2.No functional abnormalities were noted with the reservoir.The information received indicated the device had a malfunction, but because no functional abnormalities were noted with the returned components, the complaint could not be confirmed as reported.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information when attempting to inflate the device it returned to flaccid state.The device was explanted and replaced.
|
| |
|