| Model Number |
KNOTLESS T-ROPE SYN-DESMOSIS REPR KIT,SS |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a syndesmosis repair surgery the sutures broke 3 times in a row.There was no harm for patient, operator or third party reported.The surgery was finished successfully and the same device was used anyway.It was not necessary to switch the surgical technique or do a second surgery.Further information revealed that the surgery finished successfully with a new device with the same part id.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used when tensioning/tightening the sutures.
|
| |
|
Search Alerts/Recalls
|