| Catalog Number |
45750 |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Dr (b)(6) reported an issue with a mini stick max 4f x 10 cm std.018 ni/tu echo 2.75" pg that occurred during a pacemaker placement procedure.During the procedure, the proximal end of the guidewire was dissected with a bovie, during access.After accessing the nitinol guidewire, the physician started to dissect with the bovie, when he came into contact with the guidewire and sliced it.This had never occurred prior, and he was questioning the material of the guidewire and why it occurred.There was no impact or harm to the patient and the procedure was able to be completed by pulling the guidewire back.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's reported complaint description of "proximal end of the guidewire was dissected with a bovie, during access" was not confirmed.No complaint sample was returned for evaluation.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.Based on the event description the root cause of the guidewire detachment is likely due to contact with the bovie device.The guidewire device is supplied to angiodynamics by the supplier/manufacturer heraeus medical.Scar004876 was sent to heraeus for dhr review of the reported lot number.Scar004876 response stated that no deviations or nonconformances related to the event description were identified in the record review.There is no evidence to confirm the issue is due to the manufacturing process.This event is considered a non-manufacturing-related issue.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: direction for use (16600601-01) is provided with this mini-stick device.The failure mode of device fracture and embolization is cautioned as potential adverse event.No sample was returned for evaluation so it cannot be determined if device was used in a manner inconsistent with labeling.Adverse events guidewire shearing, fracture or embolization.Operational instructions: 1.Prior to insertion, flush all components with saline or heparinized/saline.2.Gain percutaneous access with the 21 g (0.9 mm) vascular introducer needle.3.Advance the 0.018 inch (0.46 mm) guidewire through the needle.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.4.Withdraw the introducer needle, leaving the 0.018 inch (0.46 mm) guidewire in place.5.Advance the coaxial assembly over the 0.018 inch (0.46 mm) guidewire.6.Simultaneously remove the dilator and 0.018 inch (0.46 mm) guidewire, while holding the sheath portion of the assembly in place.7.Advance a 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) guidewire into the sheath.8.Remove the sheath, leaving the 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) wire.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|
| |
|
Search Alerts/Recalls
|