CORCYM CANADA CORP. PERCEVAL SUTURELESS AORTIC HEART VALVE; TISSUE HEART VALVE Back to Search Results
Model Number PVS25
Medical Device Problem Codes Material Too Rigid or Stiff (1544); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Thrombocytopenia (4431)
Date of Event 12/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3 other text : device remains implanted.
 
Event or Problem Description
On (b)(6)2023, a perceval valve pvs25 was implanted in a patient.Blood platelet has decreased right after the surgery and the patient required transfusion.However, the things did not go well.Reportedly, blood platelet was more than 100,000 /l before the surgery, which is under 10,000 /l now.No further information is available at this time.
 
Event or Problem Description
On (b)(6) 2023, a percival valve pvs25 was implanted in a patient.Blood platelet has decreased right after the surgery and the patient required all types of transfusion.However, the things did not go well.Reportedly, blood platelet was more than 100,000 /l before the surgery, which is under 10,000/l now.Based on the further information received, patient had no symptoms before hospitalization.During hospitalization, paigg antibodies were elevated, and patient was diagnosed with itp thrombocytopenia (which is complicated by myasthenia gravis).As such, treatment was started.Haptoglobin was measured for suspected hemolysis, and the value dropped, leading to suspicion of hemolysis due to artificial valve, but there were no problems with pvl or pressure gradient (no abnormalities in percival function).Cardiopulmonary bypass and cross-clamp time was reported as 111min and 83 min respectively.Reportedly, patient has a history of myasthenia gravis, copd asthma, diabetes, complete atrioventricular block which required ddd pacemaker implanted prior to percival operation.Patient's medications have been reported as: myasthenia gravis medicine, antihypertensive medicine, copd antiphlogistic medicine, diabetes medicine, laxatives.
 
Additional Manufacturer Narrative
Manufacturer is retrieving and reviewing the products manufacturing records.A follow up report will be provided upon completion of this review or/and receipt of any additional information.
 
Additional Manufacturer Narrative
The manufacturing and material records for the perceval heart valve, model # icv1210, s/n # (b)(6) as they pertain to the reported event, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this valve has been manufactured and controlled in accordance with the specifications for a (model # icv1210) perceval heart valve at the time of manufacture and release.Based on the information available, it is not possible to draw a definitive root cause for the reported event.However, per the document review performed, no manufacturing deficiencies were identified with the perceval valve involved in the event.Thrombocytopenia (tcp) in cardiac surgery is a very well-known phenomenon.The etiology is multi-factorial due to multiple factors including hemodilution, blood loss with volume repletion with platelet-poor fluids, platelet activation, consumption and destruction due to contact with foreign surfaces.Tcp following aortic valve replacement (avr) is also a common phenomenon.Patient factors and prolonged extracorporeal circulation all being recognized predictors of tcp following cardiac surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEVAL SUTURELESS AORTIC HEART VALVE
Common Device Name
TISSUE HEART VALVE
Manufacturer (Section D)
CORCYM CANADA CORP.
5005 north fraser way
burnaby, bc
Manufacturer (Section G)
CORCYM CANADA CORP.
5005 north fraser way
burnaby, bc
Manufacturer Contact
laura mannino
5005 north fraser way
burnaby, bc 
MDR Report Key18514982
Report Number3004478276-2024-00111
Device Sequence Number7188564
Product Code LWR
UDI-Device Identifier00896208000436
UDI-Public(01)00896208000436(240)ICV1210(17)250302
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P150011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/02/2025
Device Model NumberPVS25
Device Catalogue NumberICV1210
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2023
Supplement Date Received by Manufacturer01/22/2024
02/26/2024
Initial Report FDA Received Date01/15/2024
Supplement Report FDA Received Date02/20/2024
03/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age81 YR
Patient SexMale
Patient Weight61 KG
-
-