Model Number 720133-01
Medical Device Problem Codes Mechanical Problem (1384); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available, the cause that contributed to the reported event cannot be established; a conclusion code of cause not established was assigned to this investigation.
 
Event or Problem Description
It was reported that this artificial urinary sphincter (aus) was not working properly.Two weeks later a revision surgery was performed, upon examination a hole in the cuff was found.The device was explanted and replaced.No patient complications were reported.
 
Additional Manufacturer Narrative
Upon receipt at our quality assurance laboratory, this artificial urinary sphincter (aus) underwent a thorough analysis.The cuff was visually inspected and underwent leak testing.A pinhole was found in the cuff shell consistent with wear at a fold which led to fluid leakage.No visual damages nor abnormalities were found neither in the pump nor the balloon, and both components passed leak testing.Based on the information available and analysis results, the cuff pinhole could have caused or contributed to the reported clinical observation of mechanical issue; therefore, a conclusion code of cause traced to component failure was assigned to this investigation.
 
Event or Problem Description
It was reported that this artificial urinary sphincter (aus) was not working properly.Two weeks later a revision surgery was performed, upon examination a hole in the cuff was found.The device was explanted and replaced.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 URINARY CONTROL SYSTEM
Common Device Name
DEVICE INCONTINENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18514983
Report Number2124215-2024-01799
Device Sequence Number7188565
Product Code EZY
UDI-Device Identifier00878953005461
UDI-Public00878953005461
Combination Product (Y/N)N
Initial Reporter CountryKS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/03/2016
Device Model Number720133-01
Device Catalogue Number720133-01
Device Lot Number877179008
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/04/2024
Supplement Date Received by Manufacturer03/21/2024
Initial Report FDA Received Date01/15/2024
Supplement Report FDA Received Date04/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/09/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BALLOON: UPN: 70400024 / LOT: 873210018; BALLOON: UPN: 70400024 / LOT: 873210018; PUMP: UPN: 72400098 / LOT: 875034016; PUMP: UPN: 72400098 / LOT: 875034016
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexMale
-
-