| Model Number |
889773 |
| Medical Device Problem Code |
Failure to Calibrate (2440)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Per the manufacturer's subsidiary's service technician, the reported issue occurred during preventative maintenance of the epgs.
|
| |
|
Event or Problem Description
|
|
Upon receipt of the device, the user reported that the electronic patient gas system (epgs) oxygen (o2) sensor failed calibration.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
During laboratory analysis, the product surveillance technician (pst) observed the output of the oxygen (o2) sensor to be 9.9 millivolts (mv) which was within the expected range of 9 - 13 mv.He installed the o2 sensor into a lab use only (luo) electronic patient gas system (epgs).Calibration was successful and the output of the epgs was found to be steady and accurate.A second calibration was also successful.The sensor operated as expected.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|