| Catalog Number |
305945 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bd safetyglide syringe contained foreign matter.The following information was provided by the initial reporter: "they opened up one needle today (b)(6) 2023 and there was a greaselike substance on the needle itself.It was under the cap on the needle.They found it with one needle so far.They did document by taking photographs and will send in." additional information received on (b)(6) 23: 1.Any adverse event or serious injury reported to patient or healthcare professional? no 2.Was there a delay of, or change in, the course of treatment due to the event? no 3.Any sample or photo available for investigation? photos are available the physical needle was placed in a sharps container with other needles.4.How was the patient outcome? are there any clinical signs, health consequences or impact? no health consequences or impact, issue with needle was noticed before tb test placed and a new needle from a different lot number was used.5.How was treatment completed for customer? new needle from a different lot number was used.6.Patient status? fine 7.Any picture of this complaint yes and they are attached.8.Please explain elaborate issue? on opening of the sterile packet for the needle the rn noticed that there was a grease like substance under the needle cap.9.Date of event? (b)(6) 2023 10.Physical available/not available? not available 11.Lot number? 2136731 we have several boxes of this lot number can we get them replaced.With this issue my rns are not comfortable using needles from this lot.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: multiple photos were provided with the customer's complaint of foreign matter on the safetyglide safety mechanism.The complaint has been verified as there is a brown substance on the safety mechanism.Without a physical sample for investigation- the root cause cannot be confirmed.This substance appears to have contaminated safety mechanism after production and is not present in the packaging.If sample were sent for investigation- ftir (fourier transform infrared spectroscopy) could have been used to inspect the substance's chemical composition but no sample has been provided.The production history was analyzed and there were no issues found during quality inspections that would lead to this issue (batch#: 2136731).
|
| |
|
Event or Problem Description
|
|
Photos received.
|
| |
|
Search Alerts/Recalls
|