WARSAW ORTHOPEDICS ELEVATE¿ SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
|
Back to Search Results |
|
| Model Number |
7770728 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from multiple sources (healthcare provider, distributor) regarding an event which occurred during an unknown spinal procedure.It was reported that for the first cage, after connecting the tools to the implant, the physician noticed fragments appearing while attempting to insert the implant.Upon removal, it was discovered that the implant had been damaged. after a previous unsuccessful attempt, the physician opted for another implant.Following the same procedure as before, when attempting to insert it, the entire implant shattered. there was no patient symptom reported.There were no further complications reported regarding the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: part 7770728 , lot # 0991350w visual and optical inspection revealed both tabs/ears of the implant has been broken off and one the upper peek portion has broken from the lower half.It appears the cage has been damaged from overload and impact force during the procedure to implant the spacer.The space may have been too small for the implant.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|